Early metallosis-related failure after total knee replacement: A report of 15 cases

40Citations
Citations of this article
60Readers
Mendeley users who have this article in their library.

Abstract

Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed. ©2011 British Editorial Society of Bone and Joint Surgery.

Cite

CITATION STYLE

APA

Willis-Owen, C. A., Keene, G. C., & Oakeshott, R. D. (2011). Early metallosis-related failure after total knee replacement: A report of 15 cases. Journal of Bone and Joint Surgery - Series B, 93 B(2), 205–209. https://doi.org/10.1302/0301-620X.93B2.25150

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free