Background/Aims: As technology continues to advance for our aging population, an increasing number of deep brain stimulation (DBS) candidates will have preexisting implanted electrical devices. In this article, we discuss safe and successful DBS in a patient with Parkinson's disease (PD) and bilateral cochlear implants. Methods: A 70-year-old male with PD and bilateral cochlear implants underwent successful microelectrode-guided DBS implantation into bilateral subthalamic nuclei (STN). The patient's cochlear implant magnets were removed and replaced in the outpatient clinic for preoperative MRI and stereotactic targeting. The cochlear implants were turned off intraoperatively for STN microelectrode recordings. Results: Precise, MRI-guided stereotactic DBS implantation was possible. Intraoperative high-fidelity microelectrode recordings confirmed STN neurons with the cochlear implants turned off. These recordings were not possible with active cochlear implant devices. Our literature review describes the other approaches/techniques that have been used to manage DBS surgery in the setting of cochlear implants. Conclusions: Despite the risk of electrical interference between implanted medical devices, DBS and cochlear implants may be safe and compatible in the same patient if necessary precautions are taken.
CITATION STYLE
Buell, T. J., Ksendzovsky, A., Shah, B. B., Kesser, B. W., & Elias, W. J. (2015, August 20). Deep Brain Stimulation in the Setting of Cochlear Implants: Case Report and Literature Review. Stereotactic and Functional Neurosurgery. S. Karger AG. https://doi.org/10.1159/000380824
Mendeley helps you to discover research relevant for your work.