Patient dose survey of mammography systems in the UK in 2013–2015

2Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A nation-wide survey of patient dose data was carried out, using data recorded in 2013–2015. Data from 32,000 women were collected. The average dose for oblique views, for DR systems, was 1.65 mGy for all women, and 1.35 mGy for 50–60 mm breasts. There was a wide range of doses for different systems, with the highest more than twice the dose of the lowest (2.03 mGy and 0.91 mGy respectively for the Hologic Dimensions and Philips MicroDose L30 systems, averaged over all breasts). Image quality, as indicated by the threshold gold thickness for 0.25 mm details, was better (0.21 μm) for the Hologic systems; for all the others it was practically the same (0.28 μm), although their doses to the average breast varied over a wide range.

Author supplied keywords

Cite

CITATION STYLE

APA

Oduko, J., & Young, K. (2016). Patient dose survey of mammography systems in the UK in 2013–2015. In Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics) (Vol. 9699, pp. 327–334). Springer Verlag. https://doi.org/10.1007/978-3-319-41546-8_41