Purpose: This paper describes and experimentally validates a methodology for improving contrast and spatial resolution of the x-ray dark-field imaging (XDFI) by cutting the monochromator-collimator asymmetrically and thinning the Laue angle analyzer. Methods: We measure the spatial resolution of our XDFI setup using a test object consisting of wolfram tungsten meshes and compare it with the theoretical prediction. Using x-ray dark-field computed tomography of breast cancer specimens (lobular carcinoma in situ), we demonstrate that the resolution of XDFI is sufficient for histopathologic analysis. Results: Our experimental results show that the overall spatial resolution of XDFI can be improved by approximately a factor of 2 when these modifications are implemented. The reconstructed images of breast cancer specimens provide sufficient details for radiologic histopathology. Conclusions: By cutting the monochromator-collimator and thinning the Laue angle analyzer, XDFI can achieve the resolution sufficient for radiologic histopathology.
CITATION STYLE
Ando, M., Nakao, Y., Jin, G., Sugiyama, H., Sunaguchi, N., Sung, Y., … Gupta, R. (2020). Improving contrast and spatial resolution in crystal analyzer-based x-ray dark-field imaging: Theoretical considerations and experimental demonstration. Medical Physics, 47(11), 5505–5513. https://doi.org/10.1002/mp.14442