Development of a physical geometric phantom for deformable image registration credentialing of radiotherapy centers for a clinical trial

4Citations
Citations of this article
26Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: This study aimed to develop a physical geometric phantom for the deformable image registration (DIR) credentialing of radiotherapy centers for a clinical trial and tested the feasibility of the proposed phantom at multiple domestic and international institutions. Methods and materials: The phantom reproduced tumor shrinkage, rectum shape change, and body shrinkage using several physical phantoms with custom inserts. We tested the feasibility of the proposed phantom using 5 DIR patterns at 17 domestic and 2 international institutions (21 datasets). Eight institutions used the MIM software (MIM Software Inc, Cleveland, OH); seven used Velocity (Varian Medical Systems, Palo Alto, CA), and six used RayStation (RaySearch Laboratories, Stockholm, Sweden). The DIR accuracy was evaluated using the Dice similarity coefficient (DSC) and Hausdorff distance (HD). Results: The mean and one standard deviation (SD) values (range) of DSC were 0.909 ± 0.088 (0.434–0.984) and 0.909 ± 0.048 (0.726–0.972) for tumor and rectum proxies, respectively. The mean and one SD values (range) of the HD value were 5.02 ± 3.32 (1.53–20.35) and 5.79 ± 3.47 (1.22–21.48) (mm) for the tumor and rectum proxies, respectively. In three patterns evaluating the DIR accuracy within the entire phantom, 61.9% of the data had more than a DSC of 0.8 in both tumor and rectum proxies. In two patterns evaluating the DIR accuracy by focusing on tumor and rectum proxies, all data had more than a DSC of 0.8 in both tumor and rectum proxies. Conclusions: The wide range of DIR performance highlights the importance of optimizing the DIR process. Thus, the proposed method has considerable potential as an evaluation tool for DIR credentialing and quality assurance.

Cite

CITATION STYLE

APA

Kadoya, N., Sakulsingharoj, S., Kron, T., Yao, A., Hardcastle, N., Bergman, A., … Nakamura, M. (2021). Development of a physical geometric phantom for deformable image registration credentialing of radiotherapy centers for a clinical trial. Journal of Applied Clinical Medical Physics, 22(7), 255–265. https://doi.org/10.1002/acm2.13319

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free