Recently, chiral structured illumination microscopy has been proposed to image fluorescent chiral domains at sub-wavelength resolution. Chiral structured illumination microscopy is based on the combination of structured illumination microscopy, fluorescence-detected circular dichroism, and optical chirality engineering. Since circular dichroism of natural chiral molecules is typically weak, the differential fluorescence is also weak and can be easily buried by the noise, hampering the fidelity of the reconstructed images. In this work, we systematically study the impact of the noise on the quality and resolution of chiral domain images obtained by chiral SIM. We analytically describe the signal-to-noise ratio of the reconstructed chiral SIM image in the Fourier domain and verify our theoretical calculations with numerical demonstrations. Accordingly, we discuss the feasibility of chiral SIM in different experimental scenarios and propose possible strategies to enhance the signal-to-noise ratio for samples with weak circular dichroism.
CITATION STYLE
Huang, S.-Y., Singh, A. K., & Huang, J.-S. (2021). Signal and noise analysis for chiral structured illumination microscopy. Optics Express, 29(15), 23056. https://doi.org/10.1364/oe.425670