Hemodynamic results of aortic valvular replacement with the porcine xenograft valve

31Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

Abstract

Twenty three patients were evaluated by cardiac catheterization 2 to 12 mth following aortic valve replacement with the porcine xenograft valve. These hemodynamic studies established a mean peak to peak systolic gradient across the prosthesis of 23 mm Hg with a range of 6-58 mm Hg. The mean effective orifice area was calculated to be 1.25 cm2. The effective orifice area increased with increasing valve size from 0.99 cm2 for the 19 mm prosthesis to 1.44 cm2 for the 25 mm prosthesis. While in general the hemodynamics of the porcine xenograft valve are comparable to other available prostheses, the exceedingly small orifice areas (0.99 cm2 and 1.03 cm2) calculated for the 19 mm and 21 mm prostheses render their use inadvisable.