The implementation of latex membranes on centrifugal microfluidic platform has led to the development of advanced liquid handling features, with the exception of a proper reagent storage-releasing technique. This paper presents a new approach for reagent storage-releasing technique using bubbles of a bubblewrap with latex membrane tabs. The bubble was first filled with a liquid sample, sealed with a biocompatible paraffin wax, and then was placed in a chamber of the microfluidic disc. The chamber was then covered with a latex membrane tab similar to single microballoon pump. To release the stored liquid the membrane was manually deflected into the chamber to burst the bubble. The sealed bubbles are leak-free up to 106 kPa, and the average reagent recovery rate is 92%.
CITATION STYLE
Aeinehvand, M. M., Ibrahim, F., & Madou, M. J. (2016). A new approach for reagent Storage-Releasing on centrifugal microfluidic platforms using bubblewrap and latex membrane. In IFMBE Proceedings (Vol. 56, pp. 269–271). Springer Verlag. https://doi.org/10.1007/978-981-10-0266-3_56