Technical Note: Patient dose from kilovoltage radiographs during motion-synchronized treatments on Radixact®

11Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: Synchrony is a motion management system available on the Radixact linear accelerator that utilizes kilovoltage (kV) radiographs to track target motion and synchronize the delivery of radiation with the motion. Proper management of this imaging dose requires accurate quantification. The purpose of this work was to use Monte Carlo (MC) simulations to quantify organ-specific patient doses from these images for various patient anatomies. Methods: Point doses in water were measured per TG-61 for three beam qualities commonly used on the Radixact. The point doses were used to benchmark a model of the imaging system built using the Monte Carlo N-Particle (MCNP) transport code. Patient computed tomography (CT) datasets were obtained for 5 patients and 100 planar images were simulated for each patient. Patient dose was calculated using energy deposition mesh tallies. Results: The MCNP model was able to accurately reproduce the measured point doses, with a median dose difference of less than 1%. The median dose (D50%) to soft tissue from 100 radiographs among the 5 patient cases ranged from 2.0 to 4.6 mGy. The max dose (D1%) to soft tissue ranged from 6.2 to 31.0 mGy and the max dose to bony structures ranged from 20.2 to 71.7 mGy. These doses can be scaled to estimate total patient dose throughout many fractions. Conclusions: Patient dose is largely dependent on imaging protocol, patient size, and treatment parameters such as fractionation and gantry period. Organ doses from 100 radiographs (an approximate number for one fraction) on the Radixact are slightly less than the doses from Tomo MVCT setup images. Careful selection of clinical protocols and planning parameters can be used to minimize risk from these images.

References Powered by Scopus

The management of respiratory motion in radiation oncology report of AAPM Task Group 76

1955Citations
N/AReaders
Get full text

AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology

649Citations
N/AReaders
Get full text

Spektr: A computational tool for x-ray spectral analysis and imaging system optimization

274Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Clinical Implementation and Initial Experience of Real-Time Motion Tracking With Jaws and Multileaf Collimator During Helical Tomotherapy Delivery

20Citations
N/AReaders
Get full text

AAPM Task Group Report 306: Quality control and assurance for tomotherapy: An update to Task Group Report 148

13Citations
N/AReaders
Get full text

Commissioning and routine quality assurance of the Radixact Synchrony system

6Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Ferris, W. S., & Culberson, W. S. (2020, November 1). Technical Note: Patient dose from kilovoltage radiographs during motion-synchronized treatments on Radixact®. Medical Physics. John Wiley and Sons Ltd. https://doi.org/10.1002/mp.14461

Readers over time

‘20‘21‘22‘23‘24‘25036912

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 2

50%

Professor / Associate Prof. 1

25%

Researcher 1

25%

Readers' Discipline

Tooltip

Medicine and Dentistry 2

67%

Physics and Astronomy 1

33%

Save time finding and organizing research with Mendeley

Sign up for free
0