Accuracy, speed and repeatability of the voice assisted subjective refractor (VASR)

2Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: To compare the accuracy, speed and repeatability of the voice assisted subjective refractor (VASR) to traditional refractive methods. Methods: Fifty healthy adult subjects were examined by autorefractor, followed by subjective phoropter refinement. Subjects were then evaluated using the VASR (Vmax Vision) to obtain an objective and subjective result. Three total assessments were performed for each subject using each of the methods described. Corrected visual acuity was recorded for each eye after each procedure. The total time was measured for both the traditional and VASR refraction. Results: A comparison of the results obtained by traditional refraction and VASR revealed no statistically significant difference from the mean in equivalent sphere measurements (P=0.1383), and the datasets were highly correlated (r=0.993). The data comparisons for cylinder power and axis were similar (cylinder: P=0.6377, r=0.864) (axis: P=0.6991, r=0.738). VASR, on average, required 71 additional seconds to complete when compared to traditional phoropter refraction. In terms of repeatability, the average difference noted upon repeat of equivalent sphere power was 0.01 D for the phoropter (P=0.98) and 0.10 D for the VASR (P=0.23). For sphere power, the average difference was 0.02 D for the phoropter (P=0.55) and 0.07 D for the VASR (P=0.58). For cylinder power, the average difference was 0.02 D for the phoropter (P=0.11) and 0.03 D for the VASR (P=0.39). For all refractive methods, the differences between measurements amounted to ≤0.10 diopters, which is neither clinically nor statistically significant. Conclusion: Refractive error results obtained with the VASR were not statistically different from those achieved using traditional phoropter methods. Time elapsed for the VASR was slightly longer than a more traditional refractive sequence. The VASR demonstrated clinically and statistically significant repeatability of measurement, consistent with traditional refraction.

Cite

CITATION STYLE

APA

Kabat, A. G., Lievens, C. W., Newman, C. M., & Weber, J. (2019). Accuracy, speed and repeatability of the voice assisted subjective refractor (VASR). Clinical Ophthalmology, 13, 1807–1813. https://doi.org/10.2147/OPTH.S213294

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free