This study compared the biomechanical responses of 3 single crowns supported by 3 different implants under axial and off-axial loading. A standard implant (3.75 mm diameter, 13 mm length), a mini implant (3 mm diameter, 13 mm length), and a short-wide implant (5.7 mm diameter, 8 mm length) were embedded in epoxy resin by the aid of a surveyor to ensure their parallelism. Each implant supported a full metal crown made of Ni-Cr alloy with standardized dimensions. Strain gauges and finite element analysis (FEA) were used to measure the strain induced under axial and off-axial functional loads of 300 N. Results showed that mini implants recorded the highest microstrains, under both axial and off-axial loading. All implants showed a considerable increase in strain values under off-axial loading. Standard and short-wide implants proved to be preferable in supporting crowns, as the standard implant showed the lowest strains under axial and off-axial loading using FEA simulation, while the short-wide implant showed the lowest strains under nonaxial loading using strain gauge analysis.
CITATION STYLE
Kheiralla, L. S., & Younis, J. F. (2014). Peri-implant biomechanical responses to standard, short-wide, and mini implants supporting single crowns under axial and off-axial loading (an in vitro study). Journal of Oral Implantology, 40(1), 42–52. https://doi.org/10.1563/AAID-JOI-D-11-00102