Comparative Evaluations of Randomly Selected Four Point-of-Care Glucometer Devices in Addis Ababa, Ethiopia

3Citations
Citations of this article
31Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Point-of-care glucometer (PoCG) devices play a significant role in self-monitoring of the blood sugar level, particularly in the follow-up of high blood sugar therapeutic response. The aim of this study was to evaluate blood glucose test results performed with four randomly selected glucometers on diabetes and control subjects versus standard wet chemistry (hexokinase) methods in Addis Ababa, Ethiopia. Method: A prospective cross-sectional study was conducted on randomly selected 200 study participants (100 participants with diabetes and 100 healthy controls). Four randomly selected PoCG devices (CareSens N, DIAVUE Prudential, On Call Extra, i-QARE DS-W) were evaluated against hexokinase method and ISO 15197:2003 and ISO 15197:2013 standards. Results: The minimum and maximum blood sugar values were recorded by CareSens N (21 mg/dl) and hexokinase method (498.8 mg/dl), respectively. The mean sugar values of all PoCG devices except On Call Extra showed significant differences compared with the reference hexokinase method. Meanwhile, all four PoCG devices had strong positive relationship (>80%) with the reference method (hexokinase). On the other hand, none of the four PoCG devices fulfilled the minimum accuracy measurement set by ISO 15197:2003 and ISO 15197:2013 standards. In addition, the linear regression analysis revealed that all four selected PoCG overestimated the glucose concentrations. Conclusions: The overall evaluation of the selected four PoCG measurements were poorly correlated with standard reference method. Therefore, before introducing PoCG devices to the market, there should be a standardized evaluation platform for validation. Further similar large-scale studies on other PoCG devices also need to be undertaken.

References Powered by Scopus

ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY

3351Citations
N/AReaders
Get full text

System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN en ISO 15197

214Citations
N/AReaders
Get full text

The state of point-of-care testing: A european perspective

121Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Glucose measurement in diagnosis and therapy of diabetes

10Citations
N/AReaders
Get full text

Technologies for Diabetes Self-Monitoring: A Scoping Review and Assessment Using the REASSURED Criteria

4Citations
N/AReaders
Get full text

Comparative Study Between Blood Glucose Meters in a Maternity Center

0Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Wolde, M., Tarekegn, G., & Kebede, T. (2018). Comparative Evaluations of Randomly Selected Four Point-of-Care Glucometer Devices in Addis Ababa, Ethiopia. Journal of Diabetes Science and Technology, 12(3), 673–679. https://doi.org/10.1177/1932296817751747

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 8

57%

Researcher 4

29%

Professor / Associate Prof. 1

7%

Lecturer / Post doc 1

7%

Readers' Discipline

Tooltip

Medicine and Dentistry 5

42%

Nursing and Health Professions 3

25%

Engineering 2

17%

Biochemistry, Genetics and Molecular Bi... 2

17%

Save time finding and organizing research with Mendeley

Sign up for free