Background: CT artifacts induced by orthopedic implants can limit image quality and diagnostic yield. As a number of different strategies to reduce artifact extent exist, the aim of this study was to systematically compare ex vivo the impact of different CT metal artifact reduction (MAR) strategies on spine implants made of either standard titanium or carbon-fiber-reinforced-polyetheretherketone (CFR-PEEK). Methods: Spine surgeons fluoroscopically-guided prepared six sheep spine cadavers with pedicle screws and rods of either titanium or CFR-PEEK. Samples were subjected to single- and dual-energy (DE) CT-imaging. Different tube voltages (80, DE mixed, 120 and tin-filtered 150 kVp) at comparable radiation dose and iterative reconstruction versus monoenergetic extrapolation (ME) techniques were compared. Also, the influence of image reconstruction kernels (soft vs. bone tissue) was investigated. Qualitative (Likert scores) and quantitative parameters (attenuation changes induced by implant artifact, implant diameter and image noise) were evaluated by two independent radiologists. Artifact degree of different MAR-strategies and implant materials were compared by multiple ANOVA analysis. Results: CFR-PEEK implants induced markedly less artifacts than standard titanium implants (p
CITATION STYLE
Huber, F. A., Sprengel, K., Müller, L., Graf, L. C., Osterhoff, G., & Guggenberger, R. (2021). Comparison of different CT metal artifact reduction strategies for standard titanium and carbon‐fiber reinforced polymer implants in sheep cadavers. BMC Medical Imaging, 21(1). https://doi.org/10.1186/s12880-021-00554-y