A simple quality assurance test tool for the visual verification of light and radiation field congruent using electronic portal images device and computed radiography

13Citations
Citations of this article
48Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: The radiation field on most megavoltage radiation therapy units are shown by a light field projected through the collimator by a light source mounted inside the collimator. The light field is traditionally used for patient alignment. Hence it is imperative that the light field is congruent with the radiation field.Method: A simple quality assurance tool has been designed for rapid and simple test of the light field and radiation field using electronic portal images device (EPID) or computed radiography (CR). We tested this QA tool using Varian PortalVision and Elekta iViewGT EPID systems and Kodak CR system.Results: Both the single and double exposure techniques were evaluated, with double exposure technique providing a better visualization of the light-radiation field markers. The light and radiation congruency could be detected within 1 mm. This will satisfy the American Association of Physicists in Medicine task group report number 142 recommendation of 2 mm tolerance.Conclusion: The QA tool can be used with either an EPID or CR to provide a simple and rapid method to verify light and radiation field congruence. © 2012 Njeh et al; licensee BioMed Central Ltd.

Cite

CITATION STYLE

APA

Njeh, C. F., Caroprese, B., & Desai, P. (2012). A simple quality assurance test tool for the visual verification of light and radiation field congruent using electronic portal images device and computed radiography. Radiation Oncology, 7(1). https://doi.org/10.1186/1748-717X-7-49

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free