Characterization of US Food and Drug Administration Class I Recalls from 2018 to 2022 for Moderate- and High-Risk Medical Devices: A Cross-Sectional Study

0Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Medical device recalls are initiated in response to safety concerns. Class I (highest severity) recalls imply a reasonable likelihood of serious adverse events or death associated with device use. Recalled devices must be identified, assessed, and corrected or removed, upon which a recall can be terminated. Objective: To characterize Class I medical device recalls and corresponding recalled devices. Methods: This was a cross-sectional study of Class I recalls posted on the Food and Drug Administration’s annual log from January 1, 2018 to June 30, 2022 for moderate-risk and high-risk medical devices. Devices were categorized by therapeutic use, need for implantation, and life-sustaining designation; recalls were categorized by reason, status, and time elapsed. Results: There were 189 unique Class I medical device recalls, including 151 (79.9%) for moderate-risk and 34 (18.0%) for high-risk devices. Sixty-five (34.4%) recalls were for cardiovascular devices, 36 (19.0%) for implanted devices, and 37 (19.6%) for lifesustaining devices. The median number of device units recalled in the US per recall notice was 4620 (interquartile range [IQR], 57842,591), with 11 (5.8%) recalls associated with more than 1 million device units. Overall, 125 (66.1%) devices had multiple recalls, with a median of 4 (IQR, 3-11) recalls issued per recalled device. As of September 15, 2022, 50 (26.5%) recalls were terminated, with a median of 24 (IQR, 17.3-30.8) months elapsed between recall initiation and termination. Recalls were terminated more commonly among devices recalled once compared to those recalled multiple times (36.2% vs 19.2%; p=0.02) and for recalls that recommended discontinuing further use of affected devices compared to those that recommended device assessment and/or education of affected population (31.8% vs 18.2%; p=0.04). Conclusion: High-severity medical device recalls are common and affect millions of device units annually in the US. Recall termination takes a significant amount of time, putting patients at risk for serious safety concerns.

Cite

CITATION STYLE

APA

Mooghali, M., Ross, J. S., Kadakia, K. T., & Dhruva, S. S. (2023). Characterization of US Food and Drug Administration Class I Recalls from 2018 to 2022 for Moderate- and High-Risk Medical Devices: A Cross-Sectional Study. Medical Devices: Evidence and Research, 16, 111–122. https://doi.org/10.2147/MDER.S412802

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free