Prolonged burst as a new method for cardioverter-defibrillator testing

3Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

AimTo find out whether it is possible to anaesthetize patients safely without analgesia and sedation, using burst pacing prolonged until the patient becomes unconscious.MethodsOne hundred and four patients undergoing implantation or reimplantation of a cardioverter-defibrillator were included. Patients randomized into Group B underwent prolonged burst pacing without analgesia and sedation. Patients in Group T underwent a T-wave shock under analgesia and sedation. Blood samples for measurement of serum neuron-specific enolase were taken before surgery and 6, 24, and 48 h after the procedure.ResultsFrom the 104 patients, 51 were randomly assigned to Group B and 53 to Group T. Four patients from Group B were switched to Group T (ventricular fibrillation not induced by burst pacing). The clinical characteristics of both groups were similar. The mean total time of cardiac arrest was significantly longer in Group B (23.0 ± 4.4 s, median 22.7) vs. Group T (10.3 ± 3.0 s, median 10.0), P < 0.0001 (Mann-Whitney U-test). The effectiveness of both induction methods was similar (92.1% in Group B and 100% in Group T). The mean neuron-specific enolase levels after 6, 24, and 48 h were similar in Groups B and T (13.1 ± 6.3 and 11.6 ± 5.8 mg/L, 14.5 ± 7.5 and 13.4 ± 6.0 mg/L, and 14.9 ± 5.9 and 12.2 ± 6.0 mg/L, respectively) as were these levels compared with baseline neuron-specificenolase levels (14.0 ± 5.9 and 13.4 ± 4.0 mg/L, respectively), P = NS for all.ConclusionDespite a longer time of total cardiac arrest, prolonged burst pacing appears to be a safe and effective method for induction of ventricular fibrillation during cardioverter-defibrillator testing, which enables omission of analgesia and sedation or general anaesthesia. © The Author 2012.

Cite

CITATION STYLE

APA

Lukl, J., Marek, D., Bulava, A., Fedorco, M., Schneiderka, P., Táborský, M., & Zapletalová, J. (2013). Prolonged burst as a new method for cardioverter-defibrillator testing. Europace, 15(1), 55–59. https://doi.org/10.1093/europace/eus250

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free