To evaluate the properties of two nickel-titanium (NiTi) reciprocating endodontic instruments (commercially known as Procodile and Reziflow), a total of 40 size 25 and 0.06 taper new Procodile and Reziflow instruments (n = 20) were subjected to cyclic fatigue tests (60° angle of curvature, 5-mm radius) at 20 °C and 37 °C and a torsional test based on ISO 3630-1. The fracture surface of each fragment was examined. The morphological, mechanical, chemical, thermal, and phase composition characteristics of the files were investigated by field-emission gun scanning electron microscopy (FEG-SEM) equipped with an energy-dispersive X-ray (EDX) detector, focused ion beam analysis (FIB), micro-Raman spectroscopy, X-ray diffraction (XRD), differential scanning calorimetry (DSC), and Auger electron spectroscopy (AES). Reziflow showed higher cyclic fatigue resistance than Procodile at 37 °C (p < 0.05). The maximum torsional strength of Procodile was lower than that of Reziflow (p < 0.05). No difference was found between their angular rotations to fracture (p > 0.05). SEM, FIB, Micro-Raman, and AES analyses revealed the presence of an Nb/Nb2O5 coating on the Procodile surface. DSC and XRD analysis confirmed that both files consist of an almost austenitic phase structure at 37 °C. The cyclic fatigue resistance of Procodile and Reziflow significantly decreases upon exposure to body temperature.
CITATION STYLE
Generali, L., Malovo, A., Bolelli, G., Borghi, A., La Rosa, G. R. M., Puddu, P., … Pedullà, E. (2020). Mechanical properties and metallurgical features of new green NiTi reciprocating instruments. Materials, 13(17). https://doi.org/10.3390/MA13173736