Optimization of analyzer-based imaging systems for minimal surface absorbed dose

1Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Analyzer-based imaging has improved tissue X-ray imaging beyond what conventional radiography was able to achieve. The extent of the improvement is dependent on the crystal reflection used in the monochromator and analyzer combination, the imaging photon energy, the geometry of the sample and the imaging detector. These many factors determine the ability of the system to distinguish between various bone tissues or soft tissues with a specified statistical certainty between pixels in a counting detector before any image processing. The following discussion will detail changes in the required number of imaging photons and the resulting surface absorbed dose when the imaging variables are altered. The process whereby the optimal imaging parameters to deliver the minimum surface absorbed dose to a sample while obtaining a desired statistical certainty between sample materials for an arbitrary analyzer-based imaging system will be described. Two-component samples consisting of bone and soft tissue are discussed as an imaging test case. The two-component approach will then be generalized for a multiple-component sample. © 2013 International Union of Crystallography Printed in Singapore-all rights reserved.

Cite

CITATION STYLE

APA

Bewer, B. E. (2013). Optimization of analyzer-based imaging systems for minimal surface absorbed dose. Journal of Synchrotron Radiation, 20(3), 405–412. https://doi.org/10.1107/S090904951300232X

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free