Suboptimal handling of piccolo samples or reagent discs for consideration in ebola response

1Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

Abstract

Operating clinical analyzers within recommended parameters can be challenging during outbreak response. Using the Piccolo Xpress point-of-care blood chemistry analyzer on guinea pig blood, we found that values of many analytes are still readily comparable when samples and reagent discs are handled at various conditions outside of manufacturers recommendations.

Cite

CITATION STYLE

APA

Spengler, J. R., Welch, S. R., Genzer, S. C., Coleman-McCray, J. A., Harmon, J. R., Nichol, S. T., & Spiropoulou, C. F. (2019). Suboptimal handling of piccolo samples or reagent discs for consideration in ebola response. Emerging Infectious Diseases, 25(6), 1238–1240. https://doi.org/10.3201/eid2506.181928