Comparison of 3 Dynamic External Fixation Devices for Proximal Interphalangeal Joint Dorsal Fracture-Dislocations in a Cadaver Model

2Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: Several improvised dynamic external fixation devices are used for treating unstable dorsal proximal interphalangeal (PIP) joint fracture-dislocations. We compared the effectiveness of 3 constructs for simulated dorsal PIP joint fracture-dislocations in a cadaver model. Methods: We tested 30 digits from 10 fresh-frozen, thawed cadaver hands. We aimed to remove the palmar 50% of the base of each digit's middle phalanx (P2), simulating an unstable dorsal PIP joint fracture-dislocation. Each PIP joint was then stabilized via external fixation with either a pins–and–rubber-bands construct, pins-only construct, or tuberculin syringe-pins construct. We allocated 10 digits per fixation group. The finger tendons were secured to a computer-controlled stepper motor–driven linear actuator. Via this mechanism, all PIP joints were taken through 1,400 cycles of flexion-extension. With the PIP joint in neutral extension, we measured the P2 dorsal translation at baseline, after fixator stabilization, and after the motion protocol. Results: The actual mean P2 palmar defect created was 48% of the base. All PIP joints were unstable after creating the defect, with a mean initial P2 dorsal displacement of 3.7 mm. After application of the fixators, all PIP joint dislocations were reduced. The median residual P2 dorsal displacements were 0.0 mm for the pins–rubber bands group, 0.1 mm for the pins-only group, and 0.5 mm for the syringe-pins group. There were no cases of PIP joint redislocation after flexion-extension cycling, and the median dorsal P2 displacements were 0.0 mm for the pins–rubber bands group; 0.0 mm for the pins-only group; and 0.5 mm for the syringe-pins group. Conclusions: All 3 external fixators restored PIP joint stability following simulated dorsal fracture-dislocation, with all reductions maintained after motion testing. The syringe-pins construct had significantly greater median residual P2 dorsal displacement after the initial reduction and motion testing, which is of unclear clinical importance. Clinical relevance: This study informs surgeon decision-making when considering dynamic external fixator options for dorsal PIP joint fracture-dislocations.

Cite

CITATION STYLE

APA

Daniels, C. M., Abbasi, P., Sanghavi, K. K., Giladi, A. M., Katz, R. D., & Means, K. R. (2023). Comparison of 3 Dynamic External Fixation Devices for Proximal Interphalangeal Joint Dorsal Fracture-Dislocations in a Cadaver Model. Journal of Hand Surgery, 48(7), 736.e1-736.e7. https://doi.org/10.1016/j.jhsa.2022.01.019

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free