A headspace gas chromatography/mass spectrometry method was developed and validated in-house for the determination of furan in foods. The method of standard additions with d 4-furan as the internal standard was used to quantitate furan. The limit of detection and limit of quantitation (LOQ) values ranged from 0.2 and 0.6 ng/g, respectively, in apple juice to 0.9 and 2.9 ng/g, respectively, in peanut butter. Recoveries were obtained at 0.5, 1, 2, and 3 times the LOQ. At 1, 2, and 3 times the LOQ, the recoveries ranged from 89.4 to 108%, and the relative standard deviations ranged from 3.3 to 17.3% for all the matrixes. For apple juice, chicken broth, and infant formula, the averaged coefficients of determination from the linear regression analyses were >0.99 with each food fortified at 0.5, 1, 2, and 3 times the LOQ. The coefficients of determination were >0.99 for green beans and 0.96 for peanut butter with the foods fortified at 1, 2, and 3 times the LOQ. Within-laboratory precision was determined by comparing the amounts of furan found in 18 samples by 2 analysts on different days with different instruments. For most of the foods, the difference between the amounts found by each analyst was <18%. The method was used to conduct a survey of >300 foods. The furan levels found ranged from none detected to 174 ng/g.
CITATION STYLE
Nyman, P. J., Morehouse, K. M., McNeal, T. P., Perfetti, G. A., & Diachenko, G. W. (2006). Single-laboratory validation of a method for the determination of furan in foods by using static headspace sampling and gas chromatography/mass spectrometry. Journal of AOAC International, 89(5), 1417–1424. https://doi.org/10.1093/jaoac/89.5.1417