Objective:To quantitatively evaluate the anterior segment using anterior segment optical coherence tomography (AS-OCT) following Boston keratoprosthesis type 1.Methods:A retrospective study consisted of AS-OCT imaging at a single time point postoperatively in 52 eyes. Main outcomes measures include anatomical and functional anterior chamber depth (ACD), angle (ACA) and peripheral and proximal synechiae.Results:The mean time point of imaging was 19.3 months postoperatively. Average anatomical and functional ACD was 2.0 and 0.21 mm respectively, and mean ACA ranged from 2.5° to 6.14° in representative meridians. An average of 8.7 clock hours of angle closure was observed in the 25 eyes in which all meridians were imaged. The majority of eyes showed peripheral (86.5%) and proximal (67.3%) synechiae.Conclusions:AS-OCT is a useful tool for quantitative evaluation of anterior segment and angle after keratoprosthesis, which is otherwise poorly visible. The majority of eyes showed shallow ACD, extensive angle closure and synechiae formation. © 2013 Kang et al.
CITATION STYLE
Kang, J. J., Allemann, N., Vajaranant, T., de la Cruz, J., & Cortina, M. S. (2013). Anterior Segment Optical Coherence Tomography for the Quantitative Evaluation of the Anterior Segment Following Boston Keratoprosthesis. PLoS ONE, 8(8). https://doi.org/10.1371/journal.pone.0070673