Evaluating the accuracy of an aneroid sphygmomanometer in a clinical trial setting

30Citations
Citations of this article
69Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: The mercury sphygmomanometer, the "gold" standard for blood pressure measurements, has been gradually phased out in many institutions because of environmental concerns. Our on-going clinical trial compared the use of mercury vs. aneroid sphygmomanometers, before implementing a study-wide transition to the aneroid sphygmomanometer. Methods: The Diabetes Prevention Program Outcomes Study (DPPOS) studied the accuracy of the Welch Allyn Tycos 767-Series Mobile aneroid sphygmomanometer from 20 March 2006 to 21 June 2006 at multiple clinic centers. We compared readings from 997 participants in 24 clinic centers using both mercury and aneroid sphygmomanometers. Results: The study found no statistically significant difference for systolic blood pressure (SBP) (P > 0.05) and a small but significantly (P < 0.0001) lower (0.8 mm Hg) reading for diastolic blood pressure (DBP) using the aneroid sphygmomanometer. Regression analysis of aneroid vs. mercury showed regression lines (Y = 4.8 + 0.96X for SBP, Y = 3.1 + 0.95X for DBP) slightly but statistically significantly different from the line of equality (P < 0.001). Participants' age, sex, race/ethnicity, body mass index, blood pressure, and clinical center together explain about 8-10% of the variation of the difference between readings from the two sphygmomanometers. Based on the above result, on 1 August 2006, DPPOS clinics began the conversion from mercury to aneroid sphygmomanometers. Conclusions: The Welch Allyn Tycos 767-Series Mobile Aneroid model 7670-04 tested in this validation study can be used to replace mercury model in clinical trials. © 2009 American Journal of Hypertension, Ltd.

Cite

CITATION STYLE

APA

Ma, Y., Temprosa, M., Fowler, S., Prineas, R. J., Montez, M. G., Brown-Friday, J., … Whittington, T. (2009). Evaluating the accuracy of an aneroid sphygmomanometer in a clinical trial setting. American Journal of Hypertension, 22(3), 263–266. https://doi.org/10.1038/ajh.2008.338

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free