Design of an oximeter based on LED-LED configuration and FPGA technology

29Citations
Citations of this article
113Readers
Mendeley users who have this article in their library.

Abstract

A fully digital photoplethysmographic (PPG) sensor and actuator has been developed. The sensing circuit uses one Light Emitting Diode (LED) for emitting light into human tissue and one LED for detecting the reflectance light from human tissue. A Field Programmable Gate Array (FPGA) is used to control the LEDs and determine the PPG and Blood Oxygen Saturation (SpO2). The configurations with two LEDs and four LEDs are developed for measuring PPG signal and Blood Oxygen Saturation (SpO2). N-LEDs configuration is proposed for multichannel SpO2measurements. The approach resulted in better spectral sensitivity, increased and adjustable resolution, reduced noise, small size, low cost and low power consumption. © 2013 by the authors; licensee MDPI, Basel, Switzerland.

Cite

CITATION STYLE

APA

Stojanovic, R., & Karadaglic, D. (2013). Design of an oximeter based on LED-LED configuration and FPGA technology. Sensors (Switzerland), 13(1), 574–586. https://doi.org/10.3390/s130100574