Background and purpose The ED coil-10 Extra Soft (EDC-10 ES) is an extremely soft coil with excellent operability for embolization of cerebral aneurysms and is frequently used as a 'finishing coil'. The aim of this study was to evaluate the efficacy and safety of this coil. Methods Clinical data were analyzed retrospectively for 92 consecutive aneurysms for which at least one EDC-10 ES was used. Follow-up angiography at 6 months was available for 69 aneurysms. Initial and follow-up outcomes were evaluated using three-category occlusion grading. Factors that affect the packing density were also investigated. Results Of the 92 aneurysms, 63 (68.5%) were classified as complete occlusion, 11 (12.0%) as having a neck remnant and 18 (19.6%) with dome filling immediately after the embolization procedure. The mean ±SD packing density was 29.5±10.6%. Periprocedural complications were observed in three cases (3.3%). At 6 months the rate of completely occluded cases increased to 78.3% (54/69) and those for neck remnant and dome filling decreased to 11.8% (8/69) and 10.1% (7/69), respectively. Recanalization occurred in six aneurysms and one aneurysm was retreated during the follow-up period, giving an overall recurrence rate of 10.0% (7/70). Linear regression analysis showed a significant association of packing density with the percentage EDC-10 ES volume ( p<0.001) and with the maximum aneurysm size ( p<0.001). Conclusions Use of the EDC-10 ES permits safe and effective embolization of a cerebral aneurysm and contributes to a reduction in the rate of recanalization.
CITATION STYLE
Harada, K., & Morioka, J. (2013). Initial experience with an extremely soft bare platinum coil, ED coil-10 Extra Soft, for endovascular treatment of cerebral aneurysms. Journal of NeuroInterventional Surgery, 5(6), 577–581. https://doi.org/10.1136/neurintsurg-2012-010498
Mendeley helps you to discover research relevant for your work.