Design and evaluation of a non-contact bed-mounted sensing device for automated in-home detection of obstructive sleep apnea: A pilot study

17Citations
Citations of this article
57Readers
Mendeley users who have this article in their library.

Abstract

We conducted a pilot study to evaluate the accuracy of a custom built non-contact pressure-sensitive device in diagnosing obstructive sleep apnea (OSA) severity as an alternative to in-laboratory polysomnography (PSG) and a Type 3 in-home sleep apnea test (HSAT). Fourteen patients completed PSG sleep studies for one night with simultaneous recording from our load-cell-based sensing device in the bed. Subjects subsequently installed pressure sensors in their bed at home and recorded signals for up to four nights. Machine learning models were optimized to classify sleep apnea severity using a standardized American Academy of Sleep Medicine (AASM) scoring of the gold standard studies as reference. On a per-night basis, our model reached a correct OSA detection rate of 82.9% (sensitivity = 88.9%, specificity = 76.5%), and OSA severity classification accuracy of 74.3% (61.5% and 81.8% correctly classified in-clinic and in-home tests, respectively). There was no difference in Apnea Hypopnea Index (AHI) estimation when subjects wore HSAT sensors versus load cells (LCs) only (p-value = 0.62). Our in-home diagnostic system provides an unobtrusive method for detecting OSA with high sensitivity and may potentially be used for long-term monitoring of breathing during sleep. Further research is needed to address the lower specificity resulting from using the highest AHI from repeated samples.

References Powered by Scopus

Sleep-related breathing disorders in adults: Recommendations for syndrome definition and measurement techniques in clinical research

5067Citations
N/AReaders
Get full text

Clinical guideline for the evaluation, management and long-term care of obstructive sleep apnea in adults

2634Citations
N/AReaders
Get full text

Clinical practice guideline for diagnostic testing for adult obstructive sleep apnea: An American academy of sleep medicine clinical practice guideline

2099Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Automatic sleep disorders classification using ensemble of bagged tree based on sleep quality features

50Citations
N/AReaders
Get full text

Recognition of patient groups with sleep related disorders using bio-signal processing and deep learning

30Citations
N/AReaders
Get full text

Technologies for sleep monitoring at home: wearables and nearables

27Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Mosquera-Lopez, C., Leitschuh, J., Condon, J., Hagen, C. C., Rajhbeharrysingh, U., Hanks, C., & Jacobs, P. G. (2019). Design and evaluation of a non-contact bed-mounted sensing device for automated in-home detection of obstructive sleep apnea: A pilot study. Biosensors, 9(3). https://doi.org/10.3390/bios9030090

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 10

67%

Lecturer / Post doc 2

13%

Researcher 2

13%

Professor / Associate Prof. 1

7%

Readers' Discipline

Tooltip

Engineering 7

39%

Medicine and Dentistry 7

39%

Computer Science 2

11%

Nursing and Health Professions 2

11%

Save time finding and organizing research with Mendeley

Sign up for free