Dislocation of a previously successful XEN glaucoma implant into the anterior chamber: A case report

23Citations
Citations of this article
29Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: XEN glaucoma implant (XEN gel stent, Aquesys, Inc) is a new minimally invasive device approved for the treatment of glaucoma. Case presentation: A 45 year old female was being followed and treated for primary open angle glaucoma in our tertiary referral center. Due to failure of medical treatment in controlling the glaucoma, surgery was offered to the patient. The XEN 45 μ-fistula implant was successfully placed in both eyes and adequate intraocular pressure control was achieved for 4 months. The left eye pressure then increased and the XEN implant was found in the anterior chamber. Topical intraocular pressure lowering therapy had to be re-initiated to achieve adequate pressure control. Conclusion: We describe a new potential complication of the XEN glaucoma implant.

Author supplied keywords

Cite

CITATION STYLE

APA

Dervenis, N., Mikropoulou, A. M., Dervenis, P., & Lewis, A. (2017). Dislocation of a previously successful XEN glaucoma implant into the anterior chamber: A case report. BMC Ophthalmology, 17(1). https://doi.org/10.1186/s12886-017-0540-1