Boston scientific vessix™ renal denervation system

0Citations
Citations of this article
1Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The Vessix Renal Denervation System (Boston Scientific Corporation, Natick, MA, USA) consists of a denervation catheter and radiofrequency generator. The catheter is a 90 cm long balloon-tipped catheter. The balloon is 25 mm long. The balloon diameters depend on the vessel size (4 mm, 5 mm, 6 mm and 7 mm). Mounted on the surface of the balloon are (4-8, depending on the balloon size) bipolar radiofrequency electrodes distributed in a helical fashion emitting radiofrequency energy to a depth of a few millimeters into the renal artery. The catheter can be delivered over the wire into the renal artery via an 8F guide sheath (more common) or 9F guide catheter.

Cite

CITATION STYLE

APA

Bertog, S. C., Vaskelyte, L., Hofmann, I., Gafoor, S., & Sievert, H. (2015). Boston scientific vessixTM renal denervation system. In Renal Denervation: A New Approach to Treatment of Resistant Hypertension (pp. 145–146). Springer-Verlag London Ltd. https://doi.org/10.1007/978-1-4471-5223-1_17