Background: Artifacts fault image quality but handling several factors can affect it. This study was conducted to investigate the effect of object size on artifacts in cone-beam computed tomography systems. Methods: Five phantoms, each containing a titanium implant in a sheep bone block, were fabricated of various sizes ranging from XS to XL: The M phantom was the same size as the device’s field of view (FOV). The L and XL phantoms were 20 and 40% larger than the FOV while the S and XS phantoms were 20 and 40% smaller than FOV, respectively. Ballistic gelatin was used to fill the phantoms. Phantoms were scanned by NewTom VGI and HDXWill Q-FACE. The mean and standard deviation (SD) of gray values in each 120 ROI was obtained by OnDemand software. The contrast to noise ratio (CNR) was also calculated. Results: The gray value in S and M phantoms were more homogenous. The lowest SD value (10.20) was found in S phantom. The highest value for SD (125.16) was observed in XL phantom. The lowest (4.47) and highest (9.92) CNR were obtained in XL and S phantoms, respectively. HDXWill Q-FACE recorded a higher SD and a lower CNR than NewTom VGI (P < 0.05). Conclusion: Object dimensions of the FOV size or up to 20% smaller provided better image quality. Since the dimensions of soft tissue in most patients are larger than the selective FOV, it is recommended that in CBCT artifacts studies, an object with dimensions closer to the patient’s dimensions be used to better relate the results with the clinical condition, because the sample dimensions affect the amount of artifacts.
CITATION STYLE
Moshfeghi, M., Safi, Y., Różyło-Kalinowska, I., & Gandomi, S. (2022). Does the size of an object containing dental implant affect the expression of artifacts in cone beam computed tomography imaging? Head and Face Medicine, 18(1). https://doi.org/10.1186/s13005-022-00326-1
Mendeley helps you to discover research relevant for your work.