Objectives: To determine the effect of different levels of positive end-expiratory pressure (PEEP) on pulse pressure variation (PPV). Design: An observational study. Setting: Operating theatres of a tertiary training hospital. Subjects: Ventilated patients who required intra-arterial blood pressure monitoring. Outcome measures: PPV during different levels of PEEP. Method: Patients were anaesthetised by means of a standard technique and ventilated with a tidal volume of 9 ml/kg ideal body mass. The PPV was calculated at PEEP levels of 2, 5, 8 and 10 cmH2O. PPV was compared at the various PEEP levels. Results: PPV at a PEEP of 8 cmH2O and 10 cmH2O was significantly larger than that at 2 cmH2O (p-value < 0.001). PPV at a PEEP of 10 cmH2O was significantly larger than that at 8 cmH2O (p-value < 0.001). PPV at a PEEP of 8 cmH2O was larger than that at 5 cmH2O (p-value = 0.002). PPV at a PEEP of 2 and 5 cmH2O did not differ significantly (p-value = 0.194). Conclusion: We have demonstrated that, in patients with normal lungs, PEEP has a significant influence on PPV. PPV may be overestimated if PEEP ≥ 8 cmH2O is applied in patients who are ventilated with a tidal volume of 9 ml/kg. It is recommended that in patients with healthy lungs PPV should be measured at a standardised PEEP of ≤ 5 cmH2O. © SASA.
CITATION STYLE
Smith, F. J., Geyser, M., Schreuder, I., & Becker, P. J. (2012). The effect of positive end-expiratory pressure on pulse pressure variation. Southern African Journal of Anaesthesia and Analgesia, 18(6), 333–338. https://doi.org/10.1080/22201173.2012.10872874