Benchtop study of leakages across the Portex, TaperGuard, and Microcuff endotracheal tubes under simulated clinical conditions

8Citations
Citations of this article
28Readers
Mendeley users who have this article in their library.

Abstract

Objectives: To compare three endotracheal tubes for leakage across the cuff (microaspiration) under a comprehensive set of simulated clinical situations. These were the Mallinckrodt TaperGuard (Covidien, US) with a tapered polyvinyl chloride cuff; the KimVent Microcuff (Kimberly-Clark Health Care, US) with a cylindrical polyurethane cuff; and a conventional Portex (Smiths Medical International Ltd, UK) with a globular polyvinyl chloride cuff. Design: A benchtop experimental study. Setting and materials: A silicone cylinder serving as the model trachea was intubated with each of the three endotracheal tubes, one at a time. A total of 20 mL of water were added above the cuff and leakage measured every minute for 20 minutes under five simulated mechanical ventilation scenarios, including different positive end-expiratory pressure levels, and disconnection with and without spontaneous breathing efforts. Each scenario was studied under three cuff pressures of 10, 20 and 30 cm H2O, and then repeated with the application of a continuous suction force of 200 cm H2O, and leakage measured every minute for 3 minutes. Results: The outcome of interest was the cumulative amount of leakage. The Microcuff endotracheal tubes with an ultrathin polyurethane cuff consistently provided the best protection against microaspiration under all simulated clinical situations, followed by TaperGuard with a tapered cuff, and lastly Portex with a globular polyvinyl chloride cuff. Clinical scenarios associated with the greatest leakage were mechanical ventilation with zero positive end-expiratory pressure, circuit disconnection with spontaneous breathing efforts, application of suction, and a low cuff pressure. Conclusions: Microcuff endotracheal tubes outperformed TaperGuard and Portex endotracheal tubes in preventing microaspiration, which is one of the major mechanisms for ventilator-associated pneumonia.

Cite

CITATION STYLE

APA

Lau, A. C. W., Lam, S. M., & Yan, W. W. (2014). Benchtop study of leakages across the Portex, TaperGuard, and Microcuff endotracheal tubes under simulated clinical conditions. Hong Kong Medical Journal, 20(1), 7–15. https://doi.org/10.12809/hkmj133930

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free