On timing-optimized SiPMs for Cherenkov detection to boost low cost time-of-flight PET

10Citations
Citations of this article
22Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objective. Recent SiPM developments and improved front-end electronics have opened new doors in TOF-PET with a focus on prompt photon detection. For instance, the relatively high Cherenkov yield of bismuth-germanate (BGO) upon 511 keV gamma interaction has triggered a lot of interest, especially for its use in total body positron emission tomography (PET) scanners due to the crystal’s relatively low material and production costs. However, the electronic readout and timing optimization of the SiPMs still poses many questions. Lab experiments have shown the prospect of Cherenkov detection, with coincidence time resolutions (CTRs) of 200 ps FWHM achieved with small pixels, but lack system integration due to an unacceptable high power uptake of the used amplifiers. Approach. Following recent studies the most practical circuits with lower power uptake (<30 mW) have been implemented and the CTR performance with BGO of newly developed SiPMs from Fondazione Bruno Kessler tested. These novel SiPMs are optimized for highest single photon time resolution (SPTR). Main results. We achieved a best CTR FWHM of 123 ps for 2 × 2 × 3 mm3 and 243 ps for 3 × 3 × 20 mm3 BGO crystals. We further show that with these devices a CTR of 106 ps is possible using commercially available 3 × 3 × 20 mm3 LYSO:Ce,Mg crystals. To give an insight in the timing properties of these SiPMs, we measured the SPTR with black coated PbF2 of 2 × 2 × 3 mm3 size. We confirmed an SPTR of 68 ps FWHM published in literature for standard devices and show that the optimized SiPMs can improve this value to 42 ps. Pushing the SiPM bias and using 1 × 1 mm2 area devices we measured an SPTR of 28 ps FWHM. Significance. We have shown that advancements in readout electronics and SiPMs can lead to improved CTR with Cherenkov emitting crystals. Enabling time-of-flight with BGO will trigger a high interest for its use in low-cost and total-body PET scanners. Furthermore, owing to the prompt nature of Cherenkov emission, future CTR improvements are conceivable, for which a low-power electronic implementation is indispensable. In an extended discussion we will give a roadmap to best timing with prompt photons.

Cite

CITATION STYLE

APA

Gundacker, S., Borghi, G., Cherry, S. R., Gola, A., Lee, D., Merzi, S., … Kwon, S. I. (2023). On timing-optimized SiPMs for Cherenkov detection to boost low cost time-of-flight PET. Physics in Medicine and Biology, 68(16). https://doi.org/10.1088/1361-6560/ace8ee

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free