Bench test assessment of mainstream capnography during high frequency oscillatory ventilation

1Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

To assess the feasibility, stability and predictability of pCO2 measurement (PetCO2) using a main stream capnograph in a high frequency oscillatory ventilation circuit. A commercially available capnograph was mounted into a high frequency oscillatory ventilator patient circuit, adjustable CO2 flow was introduced into an artificial lung and the output of the CO2 sensor assessed under varying ventilator settings. Influence of oxygen content, pressures, heat and moisture were recorded. A linear relationship between CO2 flow rate and PetCO2 was found. Varying ventilator settings influenced the measurements, but the results for PetCO2 remained within a range of 1.5 mmHg above or under then mean measurement value. Measurements remained stable despite humidification, heat, pressure amplitudes or mean airway pressure changes. From this bench test, we conclude it is feasible to measure PetCO2 using a main stream capnograph during high frequency oscillatory conditions, these measurements were stable during the experiment. Changes in CO2 production or output can be detected. The system may prove to be of clinical value, but further in vivo measurements are warranted. © 2013 Springer Science+Business Media New York.

Cite

CITATION STYLE

APA

Hartdorff, C. M., Van Heerde, M., & Markhorst, D. G. (2014). Bench test assessment of mainstream capnography during high frequency oscillatory ventilation. Journal of Clinical Monitoring and Computing, 28(1), 63–66. https://doi.org/10.1007/s10877-013-9495-9

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free