OBJECTIVES: Determine the performance of the American Academy of Pediatrics (AAP) critical congenital heart disease (CCHD) newborn screening algorithm and the impact of an alternative algorithm. METHODS: Screening was performed on term infants without a known CCHD diagnosis at or near 24 hours of age at a tertiary birth hospital by using the AAP algorithm from 2013 to 2016. Retrospective review from the birth hospital and the area's sole pediatric cardiac center identified true-and false-positives and true-and false-negatives. A simulation study modeled the results of a modified screening algorithm with a single repeat pulse oximetry test instead of 2. RESULTS: Screening results were collected on 77 148 newborns. By using the current AAP algorithm, 77 114 (99.96%) infants passed screening, 18 infants failed for an initial saturation of <90%, and 16 failed after not attaining a passing pulse oximetry level after 3 tests. There was 1 true-positive (total anomalous pulmonary venous return), 33 falsepositives, and 6 false-negatives, yielding an overall specificity of 99.96%, a sensitivity of 14.3%, and a false-positive rate of 0.043%. Among false-positives, 10 (31.3%) had significant non-CCHD disease. Simulating the modified algorithm, sensitivity remained at 14.3%, and the false-positive rate increased to 0.054%. CONCLUSIONS: Although CCHD screening in a tertiary care birth hospital may not detect many new cases of CCHD, it can detect other important diseases in newborns. Modifying the screening algorithm to 1 repeat pulse oximetry test instead of 2 may detect additional infants with significant disease without a substantial increase in the false-positive rate.
CITATION STYLE
Diller, C. L., Kelleman, M. S., Kupke, K. G., Quary, S. C., Kochilas, L. K., & Oster, M. E. (2018). A modified algorithm for critical congenital heart disease screening using pulse oximetry. Pediatrics, 141(5). https://doi.org/10.1542/peds.2017-4065
Mendeley helps you to discover research relevant for your work.