Measurement of entrainment ratio during high frequency jet ventilation

13Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We have measured tidal (VT), entrained (Ve) and "blowback" (Vbb) volumes during high frequency jet ventilation (HFJV) through a Mallinckrodt Hi-Lo Jet tracheal tube in anaesthetized patients. The above volumes were calculated by digital integration of the appropriate regions of flow curves derived from a pneumotachograph placed between the bias flow tubing and the tracheal tube. At a driving pressure of 1 bar, lung minute ventilation increased with increasing ventilatory frequency, whilst tidal volumes decreased. The contribution of entrainment to tidal volume (Ve/VT) remained constant, although the volumes entrained were relatively small and varied widely from subject to subject. Blowback volumes were considerable, especially at ventilatory frequencies used clinically (1-2Hz). We conclude that it is not possible to entrain predictable concentrations of volatile agents from the low pressure bias flow during HFJV. © 1990 Copyright: 1990 British Journal of Anaesthesia.

Author supplied keywords

Cite

CITATION STYLE

APA

Jones, M. J., Mottram, S. D., Lin, E. S., & Smith, G. (1990). Measurement of entrainment ratio during high frequency jet ventilation. British Journal of Anaesthesia, 65(2), 197–203. https://doi.org/10.1093/bja/65.2.197