Computational fluid dynamics is an essential tool for the flow field analysis of the blood pump. The interface processing method between the dynamic/static regions will affect the accuracy of simulation results, but its influence on the simulation results is still unclear. In this study, the axial-flow blood pump was taken as the research object, and the effects of the mixing plane, frozen rotor, and sliding mesh methods on the following results were compared: flux conservation at the interface, hydraulic characteristics, and velocity field distribution. In parallel, the particle image velocimetry experiment was carried out to measure the velocity field of the impeller, the inlet, and the outlet area of the blood pump. The results show that the above methods have significant differences in flux conservation between the impeller and the back vane. The average surface energy flux’s error of frozen rotor and sliding mesh are 0.7% and 0.72%, respectively, while the mixing plane method reaches 3.6%. This nonconservative transfer affects the distribution of the downstream velocity field, and the velocity field predicted by the mixing plane at the outlet is quite different. It is suggested to use the frozen rotor method and the sliding mesh method in the simulation of the blood pump.