Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement

140Citations
Citations of this article
60Readers
Mendeley users who have this article in their library.

Abstract

Several methods of measuring the migration of the femoral component after total hip replacement have been described, but they use different reference lines, and have differing accuracies, some unproven. Statistical comparison of different studies is rarely possible. We report a study of the EBRA-FCA method (femoral component analysis using Einzel-Bild-Rontgen-Analyse) to determine its accuracy using three independent assessments, including a direct comparison with the results of roentgen stereophotogrammetric analysis (RSA). The accuracy of EBRA-FCA was better than ± 1.5 mm (95% percentile) with a Cronbach's coefficient alpha for interobserver reliability of 0.84; a very good result. The method had a specificity of 100% and a sensitivity of 78% compared with RSA for the detection of migration of over 1 mm. This is accurate enough to assess the stability of a prosthesis within a relatively limited period. The best reference line for downward migration is between the greater trochanter and the shoulder of the stem, as confirmed by two experimental analyses and a computer-assisted design.

Cite

CITATION STYLE

APA

Biedermann, R., Krismer, M., Stöckl, B., Mayrhofer, P., Ornstein, E., & Franzén, H. (1999). Accuracy of EBRA-FCA in the measurement of migration of femoral components of total hip replacement. Journal of Bone and Joint Surgery - Series B, 81(2), 266–272. https://doi.org/10.1302/0301-620X.81B2.8842

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free