Automated enzyme immunoassay for lutropin with the Abbott IMx® analyzer

1Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

Abstract

An automated enzyme immunoassay for human lutropin for use with the Abbott IMX analyzer is described. The assay provides results in approximately 40 min with a sensitivity of 0.25 int. units of LH per liter for up to 23 serum or plasma samples. Cross-reactivity with follitropin (2000 int. units/L) and thyrotropin (2 int. units/L) was negligible; it was 0.016% with human choriogonadotropin (1 x 106 int. units/L). There was no interference by high concentrations of bilirubin (0.5 g/L), hemoglobin (7.50 g/L), or triglycerides (13.5 g/L). Intra-, inter-, and total assay CVs were ≤3.75%, ≤7.1%, and ≤7.94%, respectively. Values obtained with the IMx correlated well (r = 0.98, n = 194) with values obtained with Diagnostic Products' LH Double Antibody RIA, and Serono's LH MAIAclone® assay. This assay should be useful for small to medium-size laboratories involved in the clinical diagnosis of reproductive pathology.

Cite

CITATION STYLE

APA

Ford, K. A., Baker, H. N., Baar, J. G., Balay, D., Marlewski, D., Venetucci, M., & Ogunro, E. A. (1989). Automated enzyme immunoassay for lutropin with the Abbott IMx® analyzer. Clinical Chemistry, 35(12), 2333–2335. https://doi.org/10.1093/clinchem/35.12.2333