An automated enzyme immunoassay for human lutropin for use with the Abbott IMX analyzer is described. The assay provides results in approximately 40 min with a sensitivity of 0.25 int. units of LH per liter for up to 23 serum or plasma samples. Cross-reactivity with follitropin (2000 int. units/L) and thyrotropin (2 int. units/L) was negligible; it was 0.016% with human choriogonadotropin (1 x 106 int. units/L). There was no interference by high concentrations of bilirubin (0.5 g/L), hemoglobin (7.50 g/L), or triglycerides (13.5 g/L). Intra-, inter-, and total assay CVs were ≤3.75%, ≤7.1%, and ≤7.94%, respectively. Values obtained with the IMx correlated well (r = 0.98, n = 194) with values obtained with Diagnostic Products' LH Double Antibody RIA, and Serono's LH MAIAclone® assay. This assay should be useful for small to medium-size laboratories involved in the clinical diagnosis of reproductive pathology.
CITATION STYLE
Ford, K. A., Baker, H. N., Baar, J. G., Balay, D., Marlewski, D., Venetucci, M., & Ogunro, E. A. (1989). Automated enzyme immunoassay for lutropin with the Abbott IMx® analyzer. Clinical Chemistry, 35(12), 2333–2335. https://doi.org/10.1093/clinchem/35.12.2333