Rupture of extra-corporeal circuit tubing during cardiopulmonary bypass

8Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

Abstract

Roller pumps are widely used for cardiopulmonary bypass in developing nations by virtue of proven safety during several years of institutional use and cost effectiveness. However, careful adjustment of roller occlusion is needed because they are known to cause spallation, tubing wear, and the occasional incident of rupture of tubing in the extracorporeal circuit. Rupture of polyvinylchloride tubing in the pump raceway during repair of a ventricular septal defect in a 4-year-old child is discussed. The event was managed by exclusion and replacement of the defective tubing during a short period of arrest. Use of an inappropriate boot pump and failure to detect its inclusion in the bypass circuit was a significant departure from protocol. However, because occlusion settings and duration of perfusion were within acceptable limits, a manufacturing flaw could also have contributed to tubing failure, and the event may or may not have been averted by the use of larger tubing. In conclusion, this incident reiterates the need for adherence to established protocol during assembly of the pump and draws attention to the fact that tubing integrity is not a guarantee and vigilance is warranted to handle its failure.

Cite

CITATION STYLE

APA

Sai Krishna, C., Kumar, P. V. N., Satpathy, S. K., Mohan, K. R., & Babu, V. R. (2008). Rupture of extra-corporeal circuit tubing during cardiopulmonary bypass. Journal of Extra-Corporeal Technology, 40(1), 74–76. https://doi.org/10.1051/ject/200840074

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free