Comparison of volume of blood processed on haemodialysis adequacy measurement sessions vs regular non-adequacy sessions

13Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

Abstract

Background. Knowledge that adequacy measures such as the urea reduction ratio (URR) or Kt/Vurea are being measured on haemodialysis may influence the behaviour of patients or staff such that the treatment may be better on those days. This study therefore tested the hypothesis that mean volume of blood processed (VBP), utilized as a surrogate for adequacy, is higher on adequacy measurement days than non-measurement days. Methods. Patients were identified who had been on haemodialysis over the preceding 8 months. Primary outcome was the difference in the mean VBP (in litres) on URR measurement compared with non-URR measurement days (ΔVBPU-N). Univariate and multivariate correlates of mean VBP and ΔVBPU-N were also determined. Results. Eighty-nine patients were identified who met inclusion and exclusion criteria. Linear regression demonstrated a weak relationship between VBP and URR (r = 0.24, P < 0.02). This relationship was much stronger when VBP was adjusted for patient weight (mean VBP/weight; r = 0.78, P < 0.0001). The overall mean VBP was 87.41 (± 1.2 1) and the average ΔVBPU-N was 1.1 1 (± 0.3 1) (P = 0.001). Twenty per cent of patients had a clinically relevant ΔVBPU-N of >3.6 1. Patients with a graft or fistula had a significantly higher ΔVBPU-N than patients with a tunnelled catheter. Conclusions. This study demonstrates that the average VBP is less on non-URR than on URR measurement days; this difference was clinically important in > 20% of patients. Univariate analysis indicated that the use of a fistula or graft correlated with a higher ΔVBPU-N. This implies that our current method of assessing dialysis adequacy does systematically overestimate the average delivered dose of dialysis in a subset of patients.

References Powered by Scopus

The Urea Reduction Ratio and Serum Albumin Concentration as Predictors of Mortality in Patients Undergoing Hemodialysis

1307Citations
N/AReaders
Get full text

The dose of hemodialysis and patient mortality

532Citations
N/AReaders
Get full text

Effects of Dose of Dialysis on Morbidity and Mortality

396Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Online Measurement of Urea Concentration in Spent Dialysate during Hemodialysis

44Citations
N/AReaders
Get full text

Analysis of factors associated with variability in haemodialysis adequacy

41Citations
N/AReaders
Get full text

Vascular access as a determinant of adequacy of dialysis

25Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Brimble, K. S., St Onge, J., Treleaven, D. J., & Carlisle, E. J. (2002). Comparison of volume of blood processed on haemodialysis adequacy measurement sessions vs regular non-adequacy sessions. Nephrology Dialysis Transplantation, 17(8), 1470–1474. https://doi.org/10.1093/ndt/17.8.1470

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 2

40%

Researcher 2

40%

Professor / Associate Prof. 1

20%

Readers' Discipline

Tooltip

Medicine and Dentistry 3

50%

Nursing and Health Professions 1

17%

Earth and Planetary Sciences 1

17%

Engineering 1

17%

Save time finding and organizing research with Mendeley

Sign up for free