Automated flow-injection method for cadmium determination with pre-concentration and reagent preparation on-line

16Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

Abstract

The spectrophotometric determination of Cd(II) using a flow injection system provided with a solid-phase reactor for cadmium preconcentration and on-line reagent preparation, is described. It is based on the formation of a dithizone-Cd complex in basic medium. The calibration curve is linear between 6 and 300 μg L-1 Cd(II), with a detection limit of 5.4 μg L -1, an RSD of 3.7% (10 replicates in duplicate) and a sample frequency of 11.4 h-1. The proposed method was satisfactorily applied to the determination of Cd(II) in surface, well and drinking waters.

Author supplied keywords

Cite

CITATION STYLE

APA

Di Nezio, M. S., Palomeque, M. E., & Fernández Band, B. S. (2005). Automated flow-injection method for cadmium determination with pre-concentration and reagent preparation on-line. Quimica Nova, 28(1), 145–148. https://doi.org/10.1590/S0100-40422005000100025

Register to see more suggestions