PURPOSE: To investigate the accuracy of a new optical biometer for intraocular lens (IOL) power calculation in eyes undergoing cataract surgery. METHODS: Consecutive eyes of patients undergoing cataract surgery with the same IOL model were enrolled in a prospective cohort study. Axial length (AL) and corneal power were measured with an optical biometer based on optical low-coherence interferometry and Placido-disc corneal topography. IOL power was calculated with the Hoffer Q, Holladay 1 and SRK/T formulas. For each formula the lens constant was optimized in retrospect in order to achieve a mean prediction error (PE) of zero (difference between the predicted and the postoperative refraction). Median absolute error (MedAE) and percentage of eyes with PE ±0.50 D were calculated. RESULTS: Seventy-four eyes of 74 cataract patients were enrolled. The MedAE was 0.25 D with all formulas. A PE within ±0.50 D was obtained in 89.04% of cases with the Hoffer Q and SRK/T formulas, and in 87.67% of cases with the Holladay 1 formula. CONCLUSIONS: The optical biometer investigated in the present study provides accurate measurements for IOL power calculation.
CITATION STYLE
Savini, G., Hoffer, K. J., Barboni, P., Balducci, N., Schiano-Lomoriello, D., & Ducoli, P. (2017). Correction: Accuracy of optical biometry combined with Placido disc corneal topography for intraocular lens power calculation. PLOS ONE, 12(3), e0175145. https://doi.org/10.1371/journal.pone.0175145