Performance of a Nonelectric Infant Warmer in Rwandan Health Centers

7Citations
Citations of this article
56Readers
Mendeley users who have this article in their library.

Abstract

Background. Neonatal hypothermia remains a challenge in resource-limited settings. Methods. We conducted a prospective mixed-methods cohort study in rural Rwandan health centers to assess the performance of an infant warmer we designed for low-resource settings. All hypothermic infants were eligible for enrollment. Outcomes. Safety: incidence of adverse reactions. Effectiveness: attainment of euthermia, rate of temperature rise. Feasibility: correct use of warmer, signs of wear. Interviews of caregivers and nurses. Findings. Of 102 encounters, there were no adverse reactions. Of 80 encounters for hypothermia when infants on warmer for ≥1 hour, 79 achieved euthermia; 73 in ≤2 hours. Of the 80 encounters, 64 had temperature rise ≥0.5°C/h. Of the 102 encounters, there were no instances of the warmer being prepared, used, or cleaned incorrectly. Five out of the 12 warmers exhibited wear. Interview participants were predominantly positive; some found time for readiness of warmer challenging. Interpretation. The warmer performed well. It is appropriate to study in larger scale.

Cite

CITATION STYLE

APA

May, L., Nshimyiryo, A., Kubwimana, M., Nahimana, E., Schoen, N., Gadgil, A., … Hansen, A. (2019). Performance of a Nonelectric Infant Warmer in Rwandan Health Centers. Global Pediatric Health, 6. https://doi.org/10.1177/2333794X19884820