Calibration of three capnographs for use with helium and oxygen gas mixtures

7Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Capnography is considered essential in the management of mechanically-ventilated patients. Helium, as an adjunct to mechanical ventilation, is the subject of renewed interest and used increasingly. However, helium affects the performance of infrared capnometry. We constructed a simple device to generate variable mixtures of helium, oxygen and carbon dioxide within the normal physiological range, and tested the performance of two side-stream and one in-line capnographs. We found that addition of helium to the gas mixture caused all three capnographs to underestimate the concentration of carbon dioxide. The underestimation increased as the proportion of helium increased. The maximum underestimation (30%) occurred in a 79:21 helium/oxygen mixture.

Author supplied keywords

Cite

CITATION STYLE

APA

Ball, J. A. S., & Grounds, R. M. (2003). Calibration of three capnographs for use with helium and oxygen gas mixtures. Anaesthesia, 58(2), 156–160. https://doi.org/10.1046/j.1365-2044.2003.02963.x