Oesophageal Doppler and calibrated pulse contour analysis are not interchangeable within a goal-directed haemodynamic algorithm in major gynaecological surgery

15Citations
Citations of this article
27Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background Evidence for the benefit of an intraoperative use of a goal-directed haemodynamic management has grown. We compared the oesophageal Doppler monitor (ODM, CardioQ-ODM™) with a calibrated pulse contour analysis (PCA, PiCCO2™) with regard to assessment of stroke volume (SV) changes after volume administration within a goal-directed haemodynamic algorithm during non-cardiac surgery. Methods The data were obtained prospectively in patients with metastatic ovarian carcinoma undergoing cytoreductive surgery. During surgery, fluid challenges were performed as indicated by the goal-directed haemodynamic algorithm guided by the ODM. Monitors were compared regarding precision and trending. Clinical characteristics associated with trending were studied by extended regression analysis. Results A total of 762 fluid challenges were performed in 41 patients resulting in 1524 paired measurements. The precision of ODM and PCA was 5.7% and 6.0% (P=0.80), respectively. Polar plot analysis revealed a poor trending between ODM and PCA with an angular bias of -7.1°, radial limits of agreement of -58.1° to 43.8°, and an angular concordance rate of 67.8%. Dose of norepinephrine (NE) (scaled 0.1 μg kg-1min-1) [adjusted odds ratio (OR) 0.606 (95% confidence interval, CI: 0.404-0.910); P=0.016] and changes in mean arterial pressure (MAP) to a fluid challenge (scaled 10%) [adjusted OR 0.733 (95% CI: 0.635-0.845); P<0.001] were associated with trending between ODM and PCA, whereas there was no relation to type of i.v. solution. Conclusions Despite a similar precision, ODM and PCA were not interchangeable with regard to measuring SV changes within a goal-directed haemodynamic algorithm. A decrease in interchangeability coincided with increasing NE levels and greater changes of MAP to a fluid challenge.

Cite

CITATION STYLE

APA

Feldheiser, A., Hunsicker, O., Krebbel, H., Weimann, K., Kaufner, L., Werneck, K. D., & Spies, C. (2014). Oesophageal Doppler and calibrated pulse contour analysis are not interchangeable within a goal-directed haemodynamic algorithm in major gynaecological surgery. British Journal of Anaesthesia, 113(5), 822–831. https://doi.org/10.1093/bja/aeu241

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free