Abstract

Background. Transcutaneous measurement of carbon dioxide (Tcco2) provides a non-invasive estimation of arterial carbon dioxide (Paco2). Nasal capnography (Pe'co2) is used to assess ventilation during monitored anaesthesia care (MAC) with sedation since it can readily detect apnoea. We compared the agreement between Tcco2 and Pe'co2 with Paco2 and the ability to detect hypercarbia in patients under deep sedation. Methods. Forty healthy female subjects receiving deep sedation for hysteroscopy were studied. A Tcco2 (TOSCA 500, Radiometer, Inc., Westlake, OH, USA) electrode was applied to the earlobe and Pe'co2 capnography was monitored using nasal side-stream sampling. All subjects received oxygen (3 litre min-1). Subjects were evaluated at intervals using a modified Ramsay sedation score until they reached a score ≥5. Arterial blood gas values were compared with Tcco2 and Pe'co2 values. Bland-Altman, linear regression, and receiver operator characteristics analysis were performed.Results. The mean (sd) absolute difference between the Tc co2, Pe'co2, and the Paco2 were 0.43 (0.35) and 1.06 (0.8) kPa, respectively (P=0.002). Tcco2demonstrated a mean bias (2 sd) of 0.23 (0.07-0.4) kPa with Paco2 compared with -0.93 (-1.24 to -0.63) kPa for Pe'co2. One minute before blood sampling, the sensitivity of the Tcco2 monitor for detecting Paco2 >6.65 kPa was greater than for Pe'co2 (66.7 vs 33.3, P<0.01).Conclusions. Tcco2 demonstrated better agreement with Paco2 than Pe'co2 for patients under MAC with deep sedation. Tcco2 monitoring was more sensitive for detection of Paco2 >6.65 kPa than Pe'co2. © 2010 The Author.

References Powered by Scopus

Capnography accurately detects apnea during monitored anesthesia care

157Citations
N/AReaders
Get full text

Accuracy of capnography in nonintubated surgical patients

86Citations
N/AReaders
Get full text

A randomized, controlled trial of transcutaneous carbon dioxide monitoring during ERCP

72Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Society for Obstetric Anesthesia and Perinatology Consensus Statement: Monitoring Recommendations for Prevention and Detection of Respiratory Depression Associated With Administration of Neuraxial Morphine for Cesarean Delivery Analgesia

90Citations
N/AReaders
Get full text

Monitoring and delivery of sedation

81Citations
N/AReaders
Get full text

Validity of transcutaneous PCO<inf>2</inf> in monitoring chronic hypoventilation treated with non-invasive ventilation

60Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

De Oliveira, G. S., Ahmad, S., Fitzgerald, P. C., & McCarthy, R. J. (2010). Detection of hypoventilation during deep sedation in patients undergoing ambulatory gynaecological hysteroscopy: A comparison between transcutaneous and nasal end-tidal carbon dioxide measurements. British Journal of Anaesthesia, 104(6), 774–778. https://doi.org/10.1093/bja/aeq092

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 9

39%

Researcher 7

30%

Professor / Associate Prof. 5

22%

Lecturer / Post doc 2

9%

Readers' Discipline

Tooltip

Medicine and Dentistry 21

75%

Physics and Astronomy 3

11%

Nursing and Health Professions 2

7%

Engineering 2

7%

Save time finding and organizing research with Mendeley

Sign up for free