Investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin

25Citations
Citations of this article
53Readers
Mendeley users who have this article in their library.

Abstract

Background: Research has shown that up to 33% of pressure ulcers (PUs) acquired in hospitals result from the application of a medical device. Cervical collars (C-collars) have been implicated in causing PUs, due to the mechanical force they apply to the skin. In order to improve our understanding of collar-related PUs, the present study aimed to assess the biomechanical, biochemical, and microclimate effects of C-collar design and fitting tension. Methods: A cohort of 15 healthy volunteers was fit with two different C-collars according to the manufacturer guidelines. Two further collar tensions were also defined as loose and tight for each device. Each collar condition was applied for 15 minutes, with a 10 minute refractory period. Measurements at the device–skin interface included interface pressures, inflammatory biomarkers, microclimate, range of cervical motion, and comfort scores. Results: The interface pressures at each tissue site increased monotonically with greater collar tension (p<0.01), irrespective of collar design. Biomarker analysis revealed that inflammatory cytokines (IL-1a) were elevated during collar application, with the highest increase during the tight fit condition, representing over a fourfold increase from unloaded conditions. Regardless of collar tension or type, there was an increase in temperature 1.5°C ±0.8°C compared to baseline values. Range of motion significantly decreased with greater strap tension (p<0.05), with an associated increase in discomfort. Conclusion: The present findings revealed that increasing C-collar tensions caused elevated contact pressures at the device–skin interface, with a corresponding inflammatory response at the skin. These peak contact pressures were highest at the occiput, corresponding with reported PU locations. Devices should be designed to uniformly distribute pressures, and appropriate guidance is needed for their application.

Cite

CITATION STYLE

APA

Worsley, P. R., Stanger, N. D., Horrell, A. K., & Bader, D. L. (2018). Investigating the effects of cervical collar design and fit on the biomechanical and biomarker reaction at the skin. Medical Devices: Evidence and Research, 11, 87–94. https://doi.org/10.2147/MDER.S149419

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free