Accurate on-line estimation of delivered dialysis dose by dialysis adequacy monitor (DIAMON)

0Citations
Citations of this article
2Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The aim of this study was to compare equilibrated urea Kt/V (eKt/V) from the slope of the logarithmic online UV-absorbance measurements measured by the Dialysis Adequacy Monitor (DIAMON), by eKt/V obtained from a new algorithm that can be implemented into the DIAMON prototype. As the reference the urea eKt/V was utilized obtained from the blood samples according to the rate adjustment method. The mean value of equilibrated Kt/V obtained with UV-absorbance (eKt/Va) was 1.06 ± 0.21, using the new algorithm (eKt/Vn) was 1.09 ± 0.18, and eKt/V from blood-urea (eKt/Vb) 1.09 ± 0.20 (N = 21). The mean values of eKt/V were not statistically different comparing different methods. However, both the systematic and the random error were diminished by the new algorithm. The systematic error was decreased from 2.19% to 0.312%, and the random error was decreased from 14.37% to 7.30% using the new algorithm. In summary, the DIAMON prototype can accurately and on-line estimate the dialysis dose.