This work presents a calibration of a commercial condensation particle counter using National Institute of Standards and Technology (NIST) traceable methods. By the nature of the metrology involved, this work also compares the measurement results of three independent techniques for measuring aerosol concentration: continuous flow condensation particle counter (CPC); aerosol electrometer (AE); and the aerosol concentration derived from microscopic particle counting. Because of the transient nature of aerosol, there are no concentration artifact standards such as exist for particle diameter standards.We employ a mobility classifier to produce a nearly monodisperse, 80 nm, polystyrene latex aerosol. The test aerosol is used as a challenge for the CPC and the AE, and is subsequently filter sampled for electron microscopy. Our test stand design incorporates a continuous CPC aerosol concentrationmonitor to verify the aerosol stability. The CPC determines particle concentration by single particle counting at a constant sample flow rate. The AE has been calibrated to a NIST traceable current standard. The subsequent aerosol concentration measurement is obtained by determining the electrical current produced by a charged aerosol transported to the detector by a controlled aerosol flow rate. We have NIST traceability for flow rates for all methods and a methodology to calibrate the AE to NIST traceable electrical standards. The latter provides a calibration and a determination of the uncertainty in the aerosol derived current measurement. A bias in the measurements due to multiple charged particles was observed and overcome by using an electrospray aerosol generator to produce the challenge particles. This generator was able to produce aerosol concentrations over the range of 100 particles/cm 3 to 15 000 particles/cm 3 with lower number of dimer particles (≈1%). In our work, independent measurement of aerosol concentration is obtained by quantitatively collecting samples of the airborne polystyrene latex spheres on a small pore filtermaterial and determining the number of particles collected by electron microscopy. Electron micrograph images obtained using a field-emission scanning electron microscope are analyzed using particle counting.We found the relative uncertainty in the aerosol electrometer measurements to be in excess of100%for particle concentrations of approximately 120 particles/cm 3 and approximately 5% for concentrations above 6000 particles/cm 3. The uncertainty found by the microscopy method was approximately 3%. Copyright © American Association for Aerosol Research.
CITATION STYLE
Fletcher, R. A., Mulholland, G. W., Winchester, M. R., King, R. L., & Klinedinst, D. B. (2009). Calibration of a condensation particle counter using a NIST traceable method. Aerosol Science and Technology, 43(5), 425–441. https://doi.org/10.1080/02786820802716735
Mendeley helps you to discover research relevant for your work.