The effect of 5 fracture fixation methods on bone mineral density (BMD) measurement of femurs, using dual-energy X-ray absorptiometry (DXA) was determined in a canine model. Six regions of interest were measured, including the entire femur, the diaphysis of the femur, and small regions centered over the middiaphysis of the bone (lateral middiaphyseal cortex, medial middiaphyseal cortex, middiaphyseal medullary canal, and total middiaphysis). Eight unpaired femurs were collected and scanned by use of DXA before (5 separate scans/femur) and after (5 separate scans/femur) fixation by use of 1 of 5 fixation methods. These fixation methods included: intramedullary (IM) nail; IM nail and cerclage wires; IM nail and external skeletal fixation; locked IM nail; and a dynamic compression plate (DCP). All implants were made of stainless steel. The IM nail fixation devices caused significant decreases in the DXA measurement of BMD in the small regions of interest, compared with femurs without fixation devices (mean decrease, 37.3%; P < 0.05). The locked nail caused similar, but larger, decreases in the DXA measurement of BMD, compared with the IM nail fixation methods (P < 0.05). Plate fixation caused a small, but significant (P < 0.05), decrease (2.8%) in the DXA measurement of BMD in the large regions of interest, but when all regions were averaged, it did not cause significant change in this measurement, compared with femurs without fixation devices.(ABSTRACT TRUNCATED AT 250 WORDS)
CITATION STYLE
Markel, M. D., & Bogdanske, J. J. (1994). Dual-energy X-ray absorptiometry of canine femurs with and without fracture fixation devices. American Journal of Veterinary Research, 55(6), 862–866. https://doi.org/10.2460/ajvr.1994.55.06.862
Mendeley helps you to discover research relevant for your work.